ASTM E74 PDF

This standard is issued under the fixed designation E74; the number immediately 3 For referenced ASTM standards, visit the ASTM website, , or. 17 Oct Did you know several major companies fail to follow the guidelines set fourth in ASTM E74? ASTM E74 is the standard for calibration of load. All the Controls products ASTM E74 standard. Enter now!.

Author: Malalkis Voodoolkis
Country: Lesotho
Language: English (Spanish)
Genre: History
Published (Last): 16 August 2011
Pages: 241
PDF File Size: 17.73 Mb
ePub File Size: 2.61 Mb
ISBN: 271-6-38026-982-9
Downloads: 48203
Price: Free* [*Free Regsitration Required]
Uploader: Doushicage

Henry Zumbrun is president of Morehouse Instrument Co.

CMC Workbook for ASTM E74 Uncertainty

It is not applicable for dynamic or high speed force calibrations, nor can the results of calibrations performed in accordance with these practices be assumed valid for dynamic or high speed force measurements.

Recalibration is required for a permanent zero shift exceeding w74. We recommend 20 to 30 minutes for electrical stabilization of the equipment. Historical Version s – view previous versions of standard. Follow or like us on Facebook. Quality with a Big Q.

May the Force Be With You | Quality Digest

Copyright on content held by Quality Digest or by individual authors. Apply the third series of forces Run 3.

Posted by Henry Zumbrun at 9: Using an ascending calibration curve to calibrate a device in decreasing mode can result in errors of 0. There f74 eight steps to the calibration process: Method B Deflection readings should be calculated as the difference between readings at the applied force and the average or interpolated zero force readings before and after the applied force readings.

  GRANDMAS BAG OF STORIES BY SUDHA MURTHY PDF

Warm up the instrumentation. This Week in Quality Digest Live. If the calibration is stable to within 0.

CMC Workbook for ASTM E74 Uncertainty – Cal Lab Magazine

There are two steps to the stabilization process: It is the responsibility of the user of this standard to establish e774 safety, adtm, and environmental practices and determine the applicability of regulatory limitations prior to use.

Wanna do business with a company who focuses on what matters most? Exercise the UUT two to four times. Method B deflection readings are calculated as the difference between readings at the applied force and the average or interpolated zero force readings before and after the applied force readings.

An extravaganza of industry coverage on Manufacturing Day Contact Quality Digest for reprint information. Assessing the Health of a Measurement System. Everything we do, we believe in changing how people think about force and torque calibration. Place the UUT in the test e47. Inside Quality Digest Live for October 19, Anyone calibrating in accordance with the ASTM E74 standard should first purchase a full copy of the standard here.

The FDA wants medical device manufactures to succeed, new technologies in supply chain managment. If the UUT will be used to calibrate other devices using both increasing and decreasing force, it must be calibrated in both of these modes.

  JOHN M YARBROUGH DIGITAL LOGIC APPLICATIONS AND DESIGN PDF

Allow the unit under test UUT to come to room temperature. Be sure to re-exercise the UUT prior to any change in setup. This is size-dependent and can take 24 to 48 hours. Subscribe to Our Newsletter. With slight modification, the uncertainty sheet could be used for instruments calibrated in accordance with another test method or standard.

For tension and compression calibration, intersperse the loadings. Referenced Documents purchase separately The documents listed below are referenced within the subject standard but are not provided as part of the standard.

For Class A devices this low force must be greater than the resolution of the device multiplied by ; for Class AA devices the low force must be greater than the resolution of the device multiplied by Discuss 0 Hide Comments Comment.

High-performance, 3D metrology value accessible to all industries. Home Videos Privacy Policy. This corresponds to a Any instrument that is either modified or repaired should be recalibrated.